Troubleshooting 5
Agilent 7500 Series ICP-MS Hardware Manual 5-43
Check the pressure of argon gas tank. The following causes are inferred.
• Faulty mass flow controller (Connector came off or internal fault).
• Faulty Introduction board.
1453 Error: Carrier gas MFC zero adjustment malfunction.
The Mass flow controller has failed in zero adjustment.
Check the pressure of argon gas tank. The following causes are inferred.
• Faulty mass flow controller (Connector came off or internal fault).
• Faulty Introduction board.
1454 Error Makeup gas MFC zero adjustment malfunction.
The Mass flow controller has failed in zero adjustment.
Check the pressure of argon gas tank. The following causes are inferred.
• Faulty mass flow controller (Connector came off or internal fault).
• Faulty Introduction board.
1455 Error: Optional gas MFC zero adjustment malfunction.
The Mass flow controller has failed in zero adjustment.
Check the pressure of optional gas tank. The following causes are inferred.
• Faulty mass flow controller (Connector came off or internal fault).
• Faulty Introduction board.
1456 Error: Ar gas controller (AMFC) malfunction.
Error from AMFC gas module during calibration check of instrument.
Verify error by restarting the instrument. The following causes are inferred.
• Disconnected, poorly connected or malfunctioning cable.
• Fault in the AMFC gas module.
1518 Error: Main Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Main board is inferred as the cause.
1519 Error: XYZ MN Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or XYZMN board is inferred as the cause.
1520 Error: Introduction Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Introduction board is inferred as the
cause.