Hardware manual

Troubleshooting 5
Agilent 7500 Series ICP-MS Hardware Manual 5-39
The torch position may not be correct. Initialize the torch box. If Time out
occurs repeatedly, the XYZMN board may be faulty.
1416 Error: Torch position X- axis initialization failed.
Initializing the X- axis has failed. Now, the torch position is not correct.
Check if the limit switch of the X- axis works properly and if there is any cause
for malfunction, such as something stuck in a direction of the X- axis origin
point. If any cause is found, get rid of it to initialize again.
1417 Error: Torch position Y- axis initialization failed.
Initializing the Y- axis has failed. Now, the torch position is not correct.
Check if the limit switch of the Y- axis works properly and if there is any cause
for malfunction such as something stuck in a direction of the Y- axis origin
point. If any cause is found, get rid of it to initialize again.
1418 Error: Torch position Z- axis initialization failed.
Initializing the Z- axis has failed. Now, the torch position is not correct.
Check if the limit switch of the Z- axis works properly and if there is any cause
for malfunction such as something stuck in a direction of the Z- axis origin
point. If any cause is found, get rid of it to initialize again.
1422 Error: Torch position initialization time out occur.
The initialization of the torch position has not finished within the limit.
The initialization of the XYZ stage cannot be performed. Check all the
communication lines and the power line are connected properly. Faulty
communication lines or XYZMN board is inferred as the causes.
1430 Error: Air intake temperature is too high.
The temperature of the intake port located at the rear side of the instrument
exceeds the limit.
Check the inlet temperature. The following causes are inferred.
The temperature near the intake port increased due to the heater.
Intake port is clogged.
Faulty intake temperature sensor or the System I/O board.
1432 Error: Exhaust temperature is too high.
The temperature of exhaust duct exceeds the limit.
Check the exhaust temperature. The following causes are inferred.
A shortage of the duct flow rate.
Sampling cone is not tightened enough.
Faulty exhaust temperature sensor or the System I/O board.
1433 Error: 24V power supply malfunction.
The 24V power supply malfunction.