Hardware manual
5-38 Agilent 7500 Series ICP-MS Hardware Manual
5 Troubleshooting
• The power of the cooling water chiller is OFF.
• Cooling water valve is closed.
• Cooling water is insufficient (possibility of a water leak).
• Cooling pipes are clogged or leaked.
• Cooling water strainer is clogged.
• Cooling chiller is faulty.
1400* Error: Cooling water (RF/WC/IF) flow rate is too low.
The flow rate of the cooling water (RF/WC/IF) is below the limit.
Check the flow rate of the cooling water. The following causes are inferred.
• The power of the cooling water chiller is OFF.
• Cooling water valve is closed.
• Cooling water is insufficient. (a possibility of a water leak).
• Cooling pipes are clogged or leaked.
• Cooling water strainer is clogged.
• Cooling chiller is faulty.
1402 Error: Cooling water temperature is too high.
The temperature of the cooling water exceeds the limit.
Check the temperature of the cooling water. The following causes are inferred.
• The power of the cooling water chiller is OFF.
• The setting of the cooling water chiller is not correct or the circular is
faulty.
1410 Error: Torch position X- axis limit SW malfunction.
The X- axis limit switch is faulty.
Limit switch of either the H or L of X- axis or the XYZMN board is faulty.
1411 Error: Torch position Y- axis limit SW malfunction.
The Y- axis limit switch is faulty.
Limit switch of either the H or L of Y- axis or the XYZMN board is faulty.
1413 Error: Torch position X- axis time out occur.
Time out occurs during the torch position moving on the X- axis.
The torch position may not be correct. Initialize the torch box. If Time out
occurs repeatedly, the XYZMN board may be faulty.
1414 Error: Torch position Y- axis time out occur.
Time out occurs during the torch position moving on the Y- axis.
The torch position may not be correct. Initialize the torch box. If Time out
occurs repeatedly, the XYZMN board may be faulty.
1415 Error: Torch position Z- axis time out occur.
Time out occurs during the torch position moving on the Z- axis.