Hardware manual

5-38 Agilent 7500 Series ICP-MS Hardware Manual
5 Troubleshooting
The power of the cooling water chiller is OFF.
Cooling water valve is closed.
Cooling water is insufficient (possibility of a water leak).
Cooling pipes are clogged or leaked.
Cooling water strainer is clogged.
Cooling chiller is faulty.
1400* Error: Cooling water (RF/WC/IF) flow rate is too low.
The flow rate of the cooling water (RF/WC/IF) is below the limit.
Check the flow rate of the cooling water. The following causes are inferred.
The power of the cooling water chiller is OFF.
Cooling water valve is closed.
Cooling water is insufficient. (a possibility of a water leak).
Cooling pipes are clogged or leaked.
Cooling water strainer is clogged.
Cooling chiller is faulty.
1402 Error: Cooling water temperature is too high.
The temperature of the cooling water exceeds the limit.
Check the temperature of the cooling water. The following causes are inferred.
The power of the cooling water chiller is OFF.
The setting of the cooling water chiller is not correct or the circular is
faulty.
1410 Error: Torch position X- axis limit SW malfunction.
The X- axis limit switch is faulty.
Limit switch of either the H or L of X- axis or the XYZMN board is faulty.
1411 Error: Torch position Y- axis limit SW malfunction.
The Y- axis limit switch is faulty.
Limit switch of either the H or L of Y- axis or the XYZMN board is faulty.
1413 Error: Torch position X- axis time out occur.
Time out occurs during the torch position moving on the X- axis.
The torch position may not be correct. Initialize the torch box. If Time out
occurs repeatedly, the XYZMN board may be faulty.
1414 Error: Torch position Y- axis time out occur.
Time out occurs during the torch position moving on the Y- axis.
The torch position may not be correct. Initialize the torch box. If Time out
occurs repeatedly, the XYZMN board may be faulty.
1415 Error: Torch position Z- axis time out occur.
Time out occurs during the torch position moving on the Z- axis.