Hardware manual
Troubleshooting 5
Agilent 7500 Series ICP-MS Hardware Manual 5-13
1104* Error: Vacuum pressure (IF/BK) is too high to keep the Penning
gauge (AN) ON.
The Penning gauge cannot be turned ON because the vacuum pressure (IF/BK)
exceeds the limit or the Penning gauge was turned OFF because the vacuum
pressure rose abnormally (When the instrument is changing from Shutdown
Mode to Standby Mode).
Check the vacuum pressure (IF/BK). The following causes are inferred.
• The vent valve is not closed or the seal is incomplete.
• The cover of the vacuum chamber is not closed or the seal is incomplete.
• The power of the rotary pump is not connected or the breaker is OFF.
• The rotary pump oil level is low or the oil is dirty.
• There is a leak in the gate valve
• The vacuum tubing has failed (the tubing is off, or a leak is caused by an
incomplete seal such as a pin hole or a defective O- ring).
• There is a fault in the Pirani gauge (IF/BK) or the System I/O Board
1105 Error: Pirani gauge (BK) malfunction (Open).
There is a fault in the backing of the Pirani gauge or the System I/O Board
(When the instrument is changing from Standby Mode to Analysis Mode).
1109* Error: Vacuum pressure (BK) is too high to keep the Penning gauge
(AN) ON.
The Penning gauge cannot be turned ON because the vacuum pressure (BK)
exceeds the limit or the Penning gauge was turned OFF because the vacuum
pressure rose abnormally.
Check the backing line pressure. The following causes are inferred.
• The vent valve is not closed or the seal is incomplete.
• The cover of the vacuum chamber is not closed or the seal is incomplete.
• The power of the rotary pump is not connected or the breaker is OFF.
• The rotary pump oil level is low or the oil is dirty.
• There is a leak in the gate valve.
• The vacuum tubing has failed (the tubing is off, or a leak is caused by an
incomplete seal such as a pin hole or a defective O- ring).
• There is a fault in the backing Pirani gauge or the System I/O Board.
1109* Error: Vacuum pressure (IF/BK) is too high to keep the Penning
gauge (AN) on.
The Penning gauge cannot be turned ON because the vacuum pressure (IF/BK)
exceeds the limit or the Penning gauge was turned OFF because the vacuum
pressure rose abnormally.