Troubleshooting 5
Agilent 7500 Series ICP-MS Hardware Manual 5-7
The vacuum tubing has failed (the tubing is off, or a leak is caused by an
incomplete seal such as a pin hole or a defective O- ring.)
There is a fault in the backing Pirani gauge or the System I/O Board.
1016* Execution Error: Vacuum pressure (IF/BK) is too high to turn on the
Penning gauge (AN).
The Penning gauge cannot work because the vacuum pressure (IF/BK) exceeds
the limit. (When the instrument is changing from the Shutdown Mode to the
Standby Mode).
Check the vacuum pressure (IF/BK). The following causes are inferred.
The vent valve is not closed or the seal is incomplete.
The cover of the vacuum chamber is not closed or the seal is incomplete.
The power of the rotary pump is not connected or the breaker is OFF.
The rotary pump oil level is low or the oil is dirty.
There is a leak in the gate valve.
The vacuum tubing has failed (the tubing is off, or a leak is caused by an
incomplete seal such as a pin hole or a defective O- ring.)
There is a fault in the Pirani gauge (IF/BK) or the System I/O Board.
1017 Execution Error: Torch position X- axis is out of range.
The torch cannot be moved to the designated position because the X- axis
position of the torch is out of range or the small cover is closed.
Ensure that the X axis position of the torch is within range. If the small cover is
not opened completely, the movable sphere of the X- axis of the torch will be
smaller.
1018 Execution Error: Lens calibration must be done in Standby mode.
The lens calibration command is only available in the Standby mode. It is not in
Standby mode now. The command was not executed.
1019 Error: Lens Slope Calibration can not complete due to hardware
failure.
The lens calibration command was to be executed but the lens slope calibration
was not done due to some failure.
The following causes are inferred.
Either the large cover or the small cover is opened, or S24V is not supplied
for other reason---check the voltage of the S24V.
The lenses or the lens and the chamber are touching each other.
There is a fault in the HV driver board.
1020 Error: Lens 0 volt offset Calibration can not complete due to
hardware failure.
The lens calibration command was attempted, but the lens offset calibration
was not done due to hardware failure.