Hardware manual
Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-133
Changing the Hardware Information
There are two types of detectors, 5184- 1983 (ETP AF222) / G1833- 65420 (ETP
AF220) and G1833- 65575 (HPK). When a different type of detector is
installed, it is necessary to change the hardware information. Both types of
detector have a unique method to optimize the supplied High Voltage. It is
necessary to identify which detector is installed, then its optimization method
is automatically selected. When the detector G1833- 65420 (ETP AF220) is
replaced with 5184- 1983 (ETP AF222), it is not necessary to execute this.
Please complete the following steps, to change the hardware information:
1 Select Maintenance>>Hardware Settings on the ICP- MS Instrument
Control window. The Hardware Settings dialog box appears.
NOTE
HPK Detector needs the following hardware:
• Ethernet communication between the Agilent 7500 ICP-MS and the MassHunter
Workstation (or ChemStation).
• In the case of using ChemStation: Revision B.01.03 or later, or Patch File revision:
B.01.02.002 or later.
If your instrument has GPIB communication interface you can only install 5184-1983 (ETP
AF222) / G1833-65420 (ETP AF220).