Hardware manual

4-124 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
Changing the Hardware Information
There are two types of the Extraction- Omega Lens Assembly, one for the
7500cx and another for the 7500cs.
It is necessary to change the hardware information when a different type of
Ion Lens is installed.
To change the hardware information for the Ion Lens, complete the following
steps:
1 Select Maintenance>>Hardware Setting on the ICP- MS Instrument
Control window. The Hardware Settings dialog box appears.
2 Confirm the Ion Lens type.
This dialog box will show the following Ion Lens types.
"ce" Lens: Extraction- Omega Lens Assembly for 7500cx
"cs" Lens: Extraction- Omega Lens Assembly for 7500cs
Figure 102 Hardware Settings dialog box (example for 7500cx)