Hardware manual

Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-105
3 Remove the cell assembly screws.
Remove the 4 screws located on the bottom side of the Reaction cell
assembly to open the Reaction cell. Remove the upper block allowing the
Octopole assembly to remain on the lower block.
Figure 85 Reaction Cell Assembly Dismantled
Cell Upper Block
Cell Exit Lens
QP Focus
Octpole Assembly 2
Cell Lower Block
Cell Assembly Screws (4)
Octpole Shaft 1
Cell Entrance Lens
Cell Focus
Cell Entrance Screw
Cell Entrance Screw
NOTE
When cleaning the Reaction cell assembly, be sure to line up the components as they are
dismantled to assure they will be refitted in the same order.
The diameter of Octopole shaft 1 differs from Octopole shaft 2. You can determine the
direction of assembly by matching them up with the shoulder height of the cell upper and
lower cell block.