4-104 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
Disassembling the Reaction Cell Assembly
Disassemble the Reaction cell assembly after you have removed it from the
vacuum chamber.
1 Remove the Plate Bias wire.
Loosen the 2 Plate Bias wire relay screws from the Cell Exit lens. Remove
the 2 Plate Bias wires from the Cell Exit lens (Figure 84).
Figure 84 Plate Bias Wire
2 Loosen the Cell Entrance and Exit Lens assembly screws a quarter turn.
To avoid damage to the Octopole from force applied while dismantling the
Octopole in the next procedure, loosen the Cell Exit Lens assembly screw
(Figure 84) and the Cell Entrance Lens assembly screw (Figure 85) a
quarter turn prior to disassembly.
CAUTION
Disassemble the Reaction cell assembly in an area free of contamination and while
wearing laboratory gloves.
Cell Exit Lens Screws (2)
Cell Exit Lens
Plate Bias Wire Relay Screws (2)
Plate Bias Wires (2)
NOTE
When cleaning the Reaction Cell assembly, lay out the dismantled components in the order
of their original configuration.
NOTE
Only loosen the Cell Entrance and Exit Lens assembly screw a quarter turn. Loosening the
screws more than that could cause the components to come apart and make the task more
difficult.