Hardware manual

4-102 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
Removing the Reaction Cell Assembly
The Reaction cell assembly is located inside the vacuum chamber, behind the
gate valve assembly. (Figure 82)
Figure 82 Reaction Cell Assembly
The following procedure describes how to remove the Reaction cell assembly:
1 Set the Agilent 7500 to shutdown mode and switch off the power.
Refer to Chapter 3, “Startup, Shutdown and Status” for details.
Reaction Cell Assembly
Reaction Cell screw
CAUTION
Disassemble the Reaction cell in an area free of contamination.
WARNING
Turn off the power to the Agilent 7500 before removing the Reaction Cell assembly
from the vacuum chamber. Otherwise, you could expose yourself to the extreme
danger of coming in contact with high voltage electricity.