Hardware manual
Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-89
• QP focus (back)
• QP focus (front)
• Spacer 1
• Omega bias (back)
• Spacer 1
• Omega(+), Omega(- )
• Spacer 1
• Omega bias (front)
• Spacer 1
• base plate
• Spacer 3
• Spacer 1
• Einzel Lens 3
• Spacer 1
• Einzel Lens 2
• Spacer 1
• Einzel Lens 1
3 Tighten Screw 1 slowly.
Tighten Screw 1, then unscrew about 1 turn. Be careful not to tighten too
firmly, as it may become impossible to install other parts after this.
4 Fix Spacer 2 and Omega contact to Omega(+) and Omega(- ) in this order
and tighten them firmly with Screw 3.
5 Re- tighten Screw 1.
Tightening screws too firmly may damage the threads of the resin screw.
Do not over tighten.
6 Firmly tighten Screw 2.
NOTE
The Einzel Lens 2 and the Einzel Lens 3 are the same part.