4-68 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
4 Move the torch box assembly to its maintenance position using the
MassHunter Workstation (or ChemStation) software.
To move the torch box assembly to its maintenance position, complete the
following steps:
a Select Instrument>>Instrument Control from the ICP- MS Top window.
The Instrument Control window appears.
b Select Maintenance>>Sample Introduction.
The Sample Introduction Maintenance dialog box appears (Figure 66).
Figure 66 Sample Introduction Maintenance Dialog Box
c Click Maintenance.
The torch box moves away for easy interface access.
NOTE
Unless the torch box cover is taken off, the cones cannot be removed. When you click
Maintenance without removing the cover, click Initialize and remove the torch box cover.
Click Maintenance again.