Hardware manual

Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-59
Maintaining the ShieldTorch
Check the torch, shield plate, and bonnet according to the following criteria
and perform maintenance as necessary.
Torch
Cracked, chipped, or deformed by high temperatures
When performing analysis requiring ultra- sensitivity (ppt)
Deposits have formed in the central torch tube
Shield Bar
Extensive corrosion or high temperature deformation
Background ArO signal can not be lowered
Bonnet
Fractured, cracked, or chipped
Locating Tools and Supplies
To maintain the shield torch, the following items are needed:
Safety glasses and laboratory gloves
5% Nitric acid
Pure water
Methanol (HPLC grade)
Cotton swabs (P/N: 9300- 2574)
NOTE
Pure water used for cleaning is indicated as water that has been generated using the
ion-exchange method and a membrane filter, and has an electrical resistance of at least
18 MΩ.