Hardware manual
4-50 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
3 Reconnect the sample and gas tubing.
Use the figures of each nebulizer to identify the sample introduction hose
and gas hoses. Connect and fasten each hose onto its fitting.
a Connect the sample uptake tube.
Attach the sample tube from the uptake channel of the peristaltic pump
to the end of the nebulizer.
b Connect the carrier gas hose.
Attach the carrier gas hose to the nebulizers vertical gas connection.
c Connect the make- up gas hose.
Attach the make- up gas hose or gas plug to the end cap gas connection.
d Connect the sample transfer line.
Attach the transfer tube from the end of the torch to the spray chamber
exhaust port. Use the clamp provided.
Testing the Nebulizer and Spray Chamber
Perform the basic maintenance test explained in 'Testing after Maintenance'
later in this chapter. Check the gas pressures and flow rates in the Instrument
Control window to determine if the gas hoses are connected tightly. See
Chapter 3, “Startup, Shutdown and Status” in this manual for more
information. Also check for leaks in the sample uptake connections. Liquid
leaks are collected in the tray under the spray chamber and expelled out of the
drain hole on the left side of the tray.
CAUTION
Do not over-tighten the nuts or the quick-connect clamps. The glass will break if too
much pressure is applied.