Agilent 7500 Series ICP-MS Hardware Manual 4-43
Replacing or Refitting the Concentric Nebulizer and the End Cap
1 Reinsert the nebulizer into the end cap.
Replace the two O- rings in the hole in the center of the end cap. Install the
nebulizer holder nut on to the end cap first. At this time, do not tighten it.
Insert the nebulizer through the nebulizer holder nut and then insert the
nebulizer into the O- rings by gently applying pressure and twisting the
nebulizer. Lubricating the nebulizer and O- rings with pure water helps the
nebulizer slide into the O- rings. Insert the nebulizer into the end cap until
the tip of the nebulizer is flush with the inside surface of the end cap.
Position the carrier gas arm of the nebulizer so that it points in the same
direction as the makeup gas fitting. Tighten the nebulizer holder nut so that
it is finger tight. The tip of the nebulizer should be positioned as shown in
the following figure.
Figure 54
2 Insert the spray chamber O- ring in the end cap.
Adjust the tip of the
nebulizer postion
to this surface.
Nebulizer
holder nu
t
CAUTION
Do not install the nebulizer together with the nebulizer holder nut to the end cap. This
may damage the screw of nebulizer holder nut.