Hardware manual
Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-37
3 Place the sample uptake tube end on a clean plane vertically as shown
the following figure.
4 Slide the tube connector toward the clean plane gently so that the
edges of the sample uptake tube and the tube connector are flush with
each other. Refer to the following figure. Be careful not to handle the
tip with force, or it may become disfigured.
Figure 50
b Insert the tube connector with the sample uptake into the sample tube
hole of the Babington Nebulizer (Figure 51). Use the shallow groove side
of the attachment tool for tube connector until the edge of the tube
connector reaches the bottom of the hole. The gap between the base of
the tube connector and the edge of the hole should be 1 mm.
Figure 51
2 Reattach the O- rings.
3 Reattach the nebulizer to the end cap.
Place the tube end on the
clean plane vertically.
Slide the connector.
Tube Connector
Clean Plane
Tube Connector
Attachment Tool for
the tube connector
1 mm