Hardware manual
4-34 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
Figure 46
3 Remove the two O- rings.
4 Complete following steps to disconnect the sample uptake tube with the
tube adapter if there is a memory problem:
a Insert the groove of the attachment tool between the tube connector and
the nebulizer body. Refer the following figure.
Figure 47 Attachment Tool for the tube connector
b Push up the connector using the tool as a lever (see the following figure).
c Grasp the connector and pull it gently out of the nebulizer body. The
Teflon tubing and the tube connector are unified.
End Cap
Thumb Screws
Sample Tubing
Removal Tool
Plate
Babington Nebulizer
Connection Port for
Make-up Gas
Groove for the attachmen
t
(Shallow groove)
Groove for the attachment
(Wide groove)