Hardware manual
Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-27
Figure 44 Nebulizer, Spray Chamber and Spray Chamber Chiller
To remove and check the nebulizer and spray chamber, complete the following
steps:
1 Put the instrument in Standby mode.
See Chapter 3, “Startup, Shutdown and Status”.
2 Remove the sample transfer line.
Remove the ball joint clamp and disconnect the connector tube (Figure 44).
If it is difficult to remove them, first slide the torch away.
Top C ove r
Connector Tube
Ball Joint Clamp
Spray Chamber
Nebulizer and End Cap
Spray Chamber Chiller