4-24 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
3 Connect 3- way Connector Block, Peristaltic Pump Tubing, Internal
Standard Tubing Kit, etc.
Figure 43
NOTE
Excessive tightening will cause leakage and clogging in the connector.
To the Nebulizer
Conductive Connector
Sample PeriPump Tube
Sample PeriPump Tube
Tube Adapter
Tube Adapter
Internal Standard PeriPump Tube
Drain PeriPump Tube
Sample Tube
Internal Standard Tube
Sample Line Connector Block
NOTE
Make sure that there is no leakage at each connection for approximately 1 minute after
sample flows through it.
If there is a leak, tighten the connection an additional 1/4 turn.
If the leak persists, repeat this procedure from the beginning. Please be reminded that
clogging in the nebulizer is also a possibility. Check the nebulizer.