Hardware manual
4-22 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
To attach this kit, complete the following steps:
1 Perform the following procedure to connect the tube adapter and pump
tube.
1 Cut the pump tube diagonally and moisten with pure water.
2 Insert the moistened pump tubing into the Teflon tubing until it is fed
completely through.
3 Cut the pump tubing together with the Teflon tubing. If the tubes are not
cut at a 90- degree angle, a leakage or blockage may occur in the
connector.
4 Slide a Teflon nut, back ferrule and front ferrule on to the tubing. The
tubing should protrude between 0.5 and 1.0 mm beyond the front ferrule.
Figure 40 Attach the Nut, Back Ferrule and Front Ferrule
5 Insert the tube end in to the tube adapter as far as possible and tighten
the set screw. Tighten the screw an additional 1/3 turn after feeling some
resistance.
6 Repeat this procedure for both ends of the pump tube.
Teflon Nut
Teflon Tube
Teflon Tube
Pump Tube
Pump Tube
Diagonal Cut
Vertical Cut
Front Ferrule
Back Ferrule
0.5-1.0mm