Instruction manual
Product Specifications
21
Name and Model Number MABIS Deluxe SmartRead
®
Plus Automatic
Digital Wrist Blood Pressure Monitor,
04-251-001
Display System Digital display/LCD
Measuring Method Oscillometric
Power Source 2 AAA batteries
Measuring Range Pressure: 30 to 280 mmHg
Pulse: 40-199 beats/minute
Accuracy Pressure: ±3 mmHg; Pulse: ±4% of reading
Pressurization Automatic pressurization by pump
Deflation Automatic pressure release valve; Optimum
cuff inflation level determined by Fuzzy Logic
Memory Built-in memory enabling display of up to
60 measurements plus an average of the last
three measurements stored in the memory.
Alarm Clock Tone
Monitor will sound three times per day if activated
WHO Indicator Display indicates level of reading in
accordance with WHO standards
Automatic Power Off
Approximately 2 minutes after last button operation
Battery Life Approximately 300 measurements
Operation
Temperature 50°F – 104°F (10°C – 40°C)
Humidity 15% to 95% RH (non-condensing)
Storage
Temperature -4°F – 122°F (-20°C – 50°C)
Humidity 15% to 95% RH (non-condensing)
Weight 3.18 oz. (without batteries)
Dimensions 2-1/2” x 3-3/8” x 1”
Wrist Circumference Range Suitable for 5” – 8-1/4” (12.5 - 21 cm)
Accessories Storage case, Instruction Guidebook, two
AAA alkaline batteries, Quick Start Guides
Specifications are subject to change without notice.
#91-032-251 03/07
©2007 MABIS Healthcare
20
Care and Maintenance
Error Codes
1. Only use a soft, dry cloth to clean your blood pressure monitor.
2. Avoid using any types of liquids on the monitor or cuff.
3. Store your monitor in the provided storage case when not in use.
4. Do not store the unit where it will be exposed to direct sunlight, dust
or humidity. Avoid extreme temperatures.
5. Never disassemble the monitor or cuff.
6. Dropping or subjecting your blood pressure monitor to strong shocks
should be avoided.
Err: Pressure is higher than 330 mmHg or setup is
incomplete.
EE: 1: Body movement or talking occurred during this
reading.
2: The measurement is incomplete or abnormally low or
high values are displayed.
Err+ Appears when battery voltage is excessively low.