10. MULTIMODE AFM
SAFETY PRECAUTIONS
Ortiz Nanomechanics Lab : 13-5037
I. TIPHOLDERS :
Only use the AFM tipholders provided with the
Multimode, do not use any early-model AFM tipholders
from older AFM's, they may short out the power supply.
II. LASER :
During and prior to laser alignment, avoid looking
directly at the laser beam, the laser spot, or reflected laser
light-especially with highly reflective samples. Staring at the
laser beam can cause eye damage.
Do not manipulate highly reflective samples while the
laser in on.
Never plug in the laser unless the head is secured on
top of the base with the piezo springs.
III. FLUIDS
Do not operate the standard tipholder in fluids.
Standard tipholders have exposed electrical signal lines that
could cause a short circuit if exposed to a conducting fluid.
When imaging in fluids use extraordinary precautions
against spillage on any component of the AFM. Fluids
must not be spilled on or around the sample stage, electronic