User's Manual
Operation RWP20.50
36
2.8 Adjusting Trigger Force (Measuring Unit PP41.00 only)
NOTICE!
If there are strong vibrations in the machine or if heavy probe inserts are used, this can lead to an
increase in the trigger force to improve the measurement results.
1. Unscrew the stylus (refer to chapter 2.2).
2. Increase or decrease the trigger force with a hexagon key AF1.3 mm (refer to Fig. 17).
Fig. 17 Adjusting trigger force
3. Screw in the stylus (refer to chapter 2.2) and align the stylus to the spindle centre (refer to chapter 2.4).
4. Calibrate the probe.
Revolutions based on lowest
force setting
≈ Trigger force with 50 mm stylus in N
Z
X/Y
min. 0
2.5
0.3
1
3.5
0.4
2
4.5
0.5
3
5.5
0.6
4
6.5
0.7
5
7.5
0.9
6
8.5
1
7
9.5
1.1
8
10.5
1.2
9
11.5
1.3
max. 10
12.5
1.4
Factory setting
+
-
Hexagon key AF1.3 mm
(0227)