2.7 Setting the Activation Method/Code and Interference Safety
2.7.1 Setting Procedure
NOTICE
Risk of material damage!
Clean and dry the probe well before opening!
Do NOT blow off with compressed air!
1. Slide the battery compartment sleeve partially down, keeping the battery held in place in the touch probe.
2. Press the push-button for approx. 1 s to display the activation method or the activation code set.
Fig. 14 Entering the settings
3. Keep pressing the push-button to set the activation method or the activation code (refer also to 2.7.2, setting
level 2).
► After approx. 3 s, the display automatically jumps to the setting of the interference safety.
4. Press the push-button to set the interference safety (refer also to 2.7.2, setting level 3).
► After approx. 3 s, the display starts to blink if the push-button is not pressed again. The push-button is now
in synchronisation mode and can be synchronised with a receiver.
NOTICE
Risk of measuring deviations!
The push-button needs to be urgently recalibrated to avoid measuring deviations after the interference safety
setting has been changed!
Battery
compartment
sleeve
Battery
Push-button