User's Manual

RWP20.50 Operation
33
2.6 Measuring Unit Replacement
1. Remove the stylus or the temperature sensor respectively according to chapter 2.2 and 2.2.1.
2. Remove the battery from the touch probe according to chapter 2.3.
3. Insert the C-spanner into the bores of the measuring unit and carefully unscrew the measuring unit in a counter-
clockwise direction.
NOTICE
Risk of material damage!
When inserting the new measuring unit, ensure that the O-rings inside the touch probe housing and on the
measuring unit are properly seated!
4. Insert the new measuring unit into the touch probe and carefully screw it in by hand.
5. Insert the C-spanner into the bores of the new measuring unit, screw the measuring unit completely to the stop.
6. Mount the stylus or the temperature sensor respectively according to chapter 2.2 and 2.2.1.
7. Reinsert the battery into the touch probe according to chapter 2.3.
INFORMATION
After changing the battery, the time for starting the touch probe (reboot) is approx. 5 s!
8. When using PP41.00 or MY21.00 align the stylus to the spindle centre.
9. Calibrate the probe.
Fig. 13 Measuring unit replacement (Example: PP41.00 -> TP44.10)
Measuring unit
PP41.00
Temperature
measuring unit
TP44.10
C-spanner
(2951)