User's Manual

RWP20.50 Operation
31
2.4 Aligning Stylus to Spindle Centre
NOTICE
Risk of material damage!
Clean and dry the probe well!
Do NOT blow off with compressed air!
1. Remove the snap ring from the side of the probe.
2. Slide battery compartment sleeve downward.
3. Adjust the probe with the adjustment screws A (4x) to within less than 5 µm (Fig. 11).
4. Tighten the adjustment screws A (4x) against each other (Fig. 11).
5. Slide the battery compartment sleeve back up again and push the snap ring back on.
6. Check the alignment of the stylus.
7. Calibrate the probe.
NOTICE
Risk of material damage!
When closing the battery compartment sleeve, make sure the O-ring is properly seated!
Fig. 11 Align stylusto spindle centre
Battery compartment
sleeve
Dial gauge
(3079)
Hexagon key
AF 2 mm (1097)
A=Adjustment screw DIN913,
M4x8(AF2 mm) (4x) (1352)
M=1.5 Nm (1.1 lbf.ft)