User's Manual
RWP20.50 Operation
29
2.2.3 Changing Stylus (Measuring Unit PP41.00)
NOTICE
Risk of material damage!
The measuring mechanism is sensitive, therefore never use excessive force when screwing in the stylus!
The torque for screwing in the stylus is M=2 Nm!
When using a break shaft adapter (91.00-S-M4/M4), only screw in the stylus with a maximal torque of 1 Nm
(0.74 lbf.ft)!
1. Unscrew the stylus from the touch probe using the mounting pin.
2. Carefully screw the new stylus with mounting pin into the touch probe (Fig. 9).
3. Align the stylus to spindle centre (refer to chapter 2.4).
4. Calibrate the probe.
Fig. 9 Stylus change with measuring unit PP41.00
Mounting pin
(0885)
Stylus
M=2 Nm (1.47 lbf.ft)