User's Manual

Operation RWP20.50
28
2.2 Changing Sensor/Stylus
2.2.1 Changing Sensors (Temperature Measuring Unit TP44.10)
NOTICE
Risk of material damage!
Clean and dry the probe well!
Do NOT blow off with compressed air!
Do not touch the sensor tip!
1. Solely hold the white base of the socket and pull carefully.
2. Press sensor completely into the socket and pay attention to the orientation of the contact plugs.
Fig. 7 Sensor change with temperature measuring unit TP44.10
2.2.2 Changing Stylus (Measuring Unit MY21.00)
NOTICE
Risk of material damage!
Do not apply any turning force to the measuring unit!
During replacement, always hold the stylus connection with the spanner AF4 mm!
1. Unscrew the stylus from the touch probe using the mounting pin.
2. Carefully screw the new stylus with mounting pin into the touch probe (Fig. 8).
3. Align the stylus to spindle centre (refer to chapter 2.4).
4. Calibrate the probe.
Fig. 8 Stylus change with measuring unit MY21.00
Mounting pin
(0885)
Stylus
M=0.6 Nm (0.44 lbf.ft)
Spanner
AF4 mm
(1665)
Temperature sensor
Plug