Operating Instructions

Operation RC-R-100
32
2.7 Temperature Measurement
Temperature Measurement
Measuring Range
5 - 50°C (5-14 mA)
41 - 122°F (5-14 mA)
Resolution
0.1°C = 20 µA
0.182°F = 20 µA
Temperature Calculation:
(x mA * 5°C/mA) - 20°C = Temperature in °C
(x mA * 9°F/mA) - 4°F = Temperature in °F
Temperature Characteristic of RC-R-100
Pin 10 (Analog OUT)
RC-R-100
Temperature Probe
RWP20.50-G-TP
Output Pin 8
Temperature
0.1°C
20 µA = ∆0.1°C
20 µA = ∆0.182°F