Operating Manual Electrically heated sample probe tube series SP®, version SP 30-H... and SP35-H... Gas sampling and gas conditioning technology 2-1.9.
Dear customer, we have made up this operating manual in such a way that all necessary information about the product can be found and understood quickly and easily. Should you still have any question, please do not hesitate to contact M&C directly or go through your appointed dealer. Respective contact addresses are to be found in the annexe to this operating manual. Please also contact our homepage www.mc-techgroup.com for further information about our products.
Content 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 General information ...................................................................................................................... 4 Declaration of conformity ............................................................................................................. 4 Safety instructions ........................................................................................................................ 5 Warranty ...........................
Head Office M&C TechGroup Germany GmbH Rehhecke 79 40885 Ratingen Germany Telephone: 02102 / 935 - 0 Fax: 02102 / 935 - 111 E - mail: info@mc-techgroup.com www.mc-techgroup.com 1 GENERAL INFORMATION The product described in this operating manual has been examined before delivery and left our works in perfect condition related to safety regulations.
3 SAFETY INSTRUCTIONS Please note the following basic safety procedures when using this equipment: Read these operating instructions carefully before start-up and use of the equipment! The information and warnings given in these operating instructions must be heeded. Work on electrical equipment is only to be carried out by trained specialists as per the regulations currently in force.
5 USED TERMS AND SIGNAL INDICATIONS DANGER! This means that death, severe physical injuries and/or important material damages will occur in case the respective safety measures are not fulfilled. WARNING! This means that death, severe physical injuries and/or important material damages may occur in case the respective safety measures are not fulfilled. This means that minor physical injuries may occur in case the respective safety measures are not fulfilled.
6 APPLICATION The electrically heated M&C sample probe tube SP30/35-H... is used in extractive sampling systems to avoid cooling and condensation of the sample in the insitu tube from the sample point to the heated sample probe SP2000-H. To avoid a premature demage by cooling and condenstaion in dust loaded processes, we recommend a heated tube type SP30-H1.1-V, including an insitu pre-filter V20-2/30, heated up with the tube. 7 DESCRIPTION The electrically heated M&C sample tube SP30-H...
Figure 2 8 Dimensions SP30-H1.1-V and pre-filter temperatures TEMPERATURE CONTROLLER For operation an external temperature control has to be mounted. As temperature controller e.g. the M&C-controller 70304G (part no. 01B8451 for 230V or 01B8451a for 115V) in wall mounting enclosure is suitable. Gas sampling and gas conditioning technology 2-1.9.
9 TECHNICAL DATA Series SP® Temperature sensor Temperature controller Probe tube length L1 Sample temperature Operating temperature max. Pre-filter SP35-H1.1 Fe-CuNi external 175mm max. 550 °C 320°C Sample gas inlet connection G3/4"i DIN ISO 228/1 Dust loading Probe tube volume Sample pressure Ambient temperature Storage temperature Ready for operation Power supply Heating capacity max. 2 g/m3 380 ml/m max.
10 PREPARATION FOR INSTALLATION Select the optimal sampling point in accordance with the generally applicable guidelines or consult the competent persons. Locate the sampling point in such a way that there is adequate space for inserting and removing the probe and pay attention to the insertion length of the probe tube. Make certain that the probe is easily accessible so that you can carry out any subsequent maintenance work without trouble.
Figure 3 12 Mounting of the heated sample tube SP30-H../SP35-H.. ELECTRICAL CONNECTIONS WARNING! NOTE! When connecting the equipment, please ensure that the supply voltage is identical with the information provided on the model type plate. Attention must be paid to the requirements of IEC 364 (DIN VDE 0100) when setting high-power electrical units with nominal voltages of up to 1000 V, together with the associated standards and stipulations.
Figure 4 Electrical connection e.g. SP30-H2 with PT100 For executions with thermocouple the temperature sensor is also connected to terminals 7 and 8. 13 STARTING WARNING! WARNING! Before starting up check whether the mains power supply voltage corresponds with the information stated on the nameplate.
14 MAINTENANCE Maintenance of the heated sample tubes is mainly limited to the change of the screwed-on or heated pre-filter (SP30-H1.1-V) and the check of the sealings. For the change or cleaning of the pre-filter the gas sample probe with the heated probe tube and the pre-filter has to be dismounted. It is difficult to give any recommendations as to a particular maintenance cycle. Depending on process conditions, a meaningful maintenance cycle must be elaborated for the specific application.