2.15
Technical specications and illustrations are without
obligation, subject to modications. 12.97/06.06
Series SP
®
Type
Version SP35- .... Version SP30- ....
H1.1 H2 H1.1 H2 H1.1-V
Temperature sensor Fe-CuNi PT100 2 Fe-CuNi PT100 2 Fe-CuNi
Temperature controller optional, external
Probe tube length L1 175 mm max. 2 m max. 1 m
Sample temperature max. 550 °C
Operating temperature max. 320 °C 200 °C 320 °C 200 °C 320 °C
Pre-filter length 520 mm, ø 60 mm,
filter porosity 2µm, integrated and heated
V20-2/30
Sample gas inlet connection G3/4"i DIN ISO 228/1 pre-filter with G1 1/2"i
DIN ISO 228/1
Dust loading max. 2 g/m
3
>2 g/m
3
Probe tube volume 380 ml/m 420 ml
Sample pressure max. 5 bar g
Ambient temperature -20 °C to +80 °C
Storage temperature -30 °C to +90 °C
Ready for operation 2 h
Power supply 230V AC, (115V optional) 230/115V AC - switchable,
Heating capacity 200W 0,6m: 600W, 1m: 800W, 1,5m: 1200W, 2m: 1200W
Electrical connections terminals, max. 2,5 mm
2
, 2x PG13,5 cable gland
Electrical standard EN 61010, EN60519-1
Degree of protection IP54 EN 60529
Mounting flange DN65 PN6, Form B with 4 welded screws on both sides M 12x 40 mm
Material of parts in contact with the sample stainless steel SS316Ti, Version SP30-H1.1/HC: Hastelloy C 1.4539, SS316Ti/SS316
Technical Data
∆P and T
90
at flow: 100 200 500 1000 Nl/h
∆P Pressure loss SP30-H... length 1m < 1 < 1 < 1 < 1 mbar
∆P pressure loss SP30-H...-V
length 1m with new pre-filter V20-2/30
< 1 < 1 1,5 4 mbar
T
90
- time 14 7 < 3 < 2 sec
Pressure difference and T90-time
Version SP35-... Version SP30- ... execution:
length „L1“ 175 mm 0,6 m 1,0 m 1,5 m 2,0 m
Part number 20S9320
20S9322
02S9023
02S9026
02S9023p
02S9025
02S9092
02S9025p
02S9026
02S9093
02S9026p
02S9024
02S9094
02S9024p
H1.1
H1.1/HC
H2
Weight 4,8 kg 6,4 kg 8,0 kg 9,8 kg 11,6 kg
Part number 02S9037 02S9038 H1.1-V
Weight 9,0 kg 10,6 kg
+ 02S9036 Standard, switchable 115V AC
Part number and weight
210
120
Ø 130
20
∅ 160
14 40
M12
L1
Terminals
Ø 42,2
G3/4"
PG13,5
Temperature-Sensor
Power
Temperature of pre filter in dependence of controller
temperature at 20°C ambient temperature.
500
ø 64
210
120
ø 130
ø 160
14
40
M12
L1
Terminals
PG13,5
Temperature
Sensor
Power
Controller temperature
180°C
250°C
300°C
Temperature of pre filter
135°C
182°C
220°C
Probe tube SP30-H1.1/H2, SP35-H1.1/H2
Dimensions in mm
Probe tube with pre-filter SP30-H1.1-V
Dimensions
The corresponding temperature controller has to be ordered and mounted separatly. See leaet 2-5.1.