14 Gas sampling and gas conditioning technology 2-1.3.9-ME
Part No. 20S9045 + 20S9012
Way of test gas /R/CC = via sample gas outlet or /R/CCF = via probe filter element
Check valve /R check valve 1/4”
Opening pressure >0,7 bar
Connection 6mm tube
Maximum blow back pressure 6 bar abs.
Maximum operating temperature +185 °C
Option high performance blow back valve /BB /BBF
Part No. 20S9008 20S9006
Way of blow back gas via filter chamber via probe filter element
Check valve high performance check valve 3/8”
Opening pressure >0,7 bar
Connection 8mm tube
Maximum blow back pressure 6 bar abs.
Maximum operating temperature 185 °C
Option test gas valve / shut-off valve /C + /I
Part No. 20S9011 + 20S9009
Way of test gas via sample gas outlet with shut-off to the process
Check valve check valve 1/4”
Opening pessure >0,7 bar
Connection 6mm tube
Shut-off valve bellow-type valve with pneumatic drive
Pressure control air 3-10 bar
Connection control air 1/8” NPT i
Option second sample gas outlet /2X
Part No. 20S9015
Connection 1/4” NPT i
Option spun glass cartridge /FW
Part No. 20S9047 20S9046
Material SS316Ti, Novapress SS316Ti, Graphite
Maximum surface temperature of the gas sample probe:
The maximum surface temperature of the gas sample probe depends on the temperature of the
process medium and the options used (heating HEX5-x.08 and backflush /RS). Under no cir-
cumstances, the sample probes SP3200(V)… must exceed the maximum surface temperature
fixed by the temperature class.
The maximum surface temperatures and inferable temperature classes of the gases are shown
in the following table.