Owner's manual
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.9
* Standard, ** optional
Technical Data
Technical specications and illustrations are without
obligation, subject to modications. 10.04/11.12
Gas sample probe type SP3000
Part No. 20S5500
Weather protection shield yes
Filter housing material Stainless steel 316 / 316Ti
Sealing materials Graphite
Probe flange sealing material Graphite
Pre-filter optional, for a valid Ex-approval according to ATEX the probe SP3000 has to be operated with a pre-
filter from page 4
Sample pressure max. 0,5-6 bar abs.
Ambient temperature -20 °C to +60 °C
Permissible process gas temperature depending on the temperature class, however max. 200°C at the probe entry
Filter chamber volume 120 cm
3
Filter element, porosity F-3SS150= stainless steel*, 3 micron S-2K150= ceramic**, 2 micron,
Sample gas outlet connection 1x 1/4“ NPTi for max. 8mm-tube connectors
Connection gas outlet at option RS 6 mm Swagelok connector
Mounting flange DN65 PN6, FormB, SS316Ti* >DN or ANSI possible**
Weight 7 kg
Marking
II 1D/2GD -20°C ≤ Ta ≤ +60°C EXAM BVS 04 ATEX H 045X
Marking with option RS
II 1D/2GD -20°C ≤ Ta ≤ +60°C EXAM BVS 04 ATEX H 045X
Option back purge unit type RS RS
Part No. 20S5560 (a)
Power supply 230V 50/60Hz 9W or 115V 50/60Hz 9W (a)
Electrical connection cable 3x1mm
2
Marking
II 2GD EEx m II 135°C
Connection G1/2” at the buffer vessel
Max. back purge pressure 6 bar abs.
Volume buffer vessel 2 liters
Ambient temperature -20°C to 60°C
Option heating type HEX4 HEX4-135 HEX4-180
Part No. 20S5510 20S5520
Control self-regulating
Power supply 115V - 230V 50/60Hz
Electrical connection cable gland, terminal range 7 - 12mm, terminals max. 4mm
2
Marking
II 2G EEx em T4 / II 2D IP66 135°C EXAM
BVS 04 ATEX E 253
II 2G EEx em T3 / II 2D IP66 180°C EXAM
BVS 04 ATEX E 253
Power 400W
Case protection IP66; EN60529
Max. temperature 120 °C 160 °C
Min. temperature 90 °C 120 °C
Ambient temperature -20 °C to +60 °C
Low temperature alarm contact <60°C, 1 contact MC-NO, 230V 1,5A AC, 0,5ADC <100°C, 1 contact MC-NO, 230V 1,5A AC, 0,5ADC
Option 2-way-ball valve in the probe entrance /VA
Part No. 20S9050
Operating temperature -20°C up to +185°C
Option 2/3-way-ball valve in the probe entrance /3VA
Part No. 20S9325
Backflush / Test gas connection 6mm tube
Operating temperature -20°C up to +185°C
Option pneum. drive for ball valve /VA o. /3VA MS1
Part No. 20S9055
Connection control air G1/4” i
Pressure control air 5-10 bar
Temperature class T4
Option second sample gas outlet /2X
Part No. 20S9015
Connection 1/4” NPT i