Operator's manual User Manual
2-1.1-ME Gas sampling and gas conditioning technology 9
8 TECHNICAL DATA
Gas sample probe type SP2000 SP2000-H SP2300-H SP2400-H
Part no. 20S1000 20S2000 20S3000 20S3500
Weather protection shield no yes yes yes
Protection class terminal box IP54 EN60529
Material filter housing Stainless steel SS316/316Ti* PTFE Titanium
Sealing material FPM* /7aT** = PTFE -H320/C** = Graphite
Probe flange sealing material Novapress
Insitu probe tube / prefilter Optional
Sample pressure max. 0,4-6 bar* abs., /7aT**= 2 bar abs.,
/HP** = 25 bar abs.
2 bar abs. 0,4-6 bar abs.
Ambient temperature 20 °C to +180 °C -20 °C to +60 °C*;
/PT100, /Fe-CuNi, /Ni-CrNi** = -20°C to +80°C
Filter chamber volume 120 cm
3
Filter porosity S-2K150= ceramic*, 2micron, /F-0, 1GF150= glass fibre**, 0,1 micron,
/F-3SS150= stainless steel**, 3 micron, /FW= spun glass**
Thermostat, temperature adjust-
ment
0-180°C* -H320/C**= 0-320°C /PT100** /Fe-CuNi** /Ni-
CrNi**
Ready for operation after 40min -H320/C** = after 60min
Low temperature alarm contact* contact rating: 250V, 3A~, 0,25A= alarm point: ∆T 30°C
Sample gas outlet cnnection 1x 1/4“ NPTi* tube connection** ø 6, 8 or 10mm
-H320/C**= tube connection 6mm or 8mm*
Backflush/Test gas connection
1/4“ NPTi* /R**, -H320/C**= tube ø 6 mm
Power supply
230V 50/60 Hz, 800W /115V** = 115V 60Hz, 800W (fuse
protection 10A)
Electrical connections Terminals max. 4mm
2
, 2 x cable gland PG 13,5
Electrical equipment standard
EN 61010, EN 60519-1
Mounting flange DN65 PN6, form B >DN or ANSI possible** /HP** = DN50 PN25
Mounting flange material SS316Ti PTFE Titanium
Weight 7 kg* 15,4 kg* 15,4 kg* 14,5kg*
* = Standard
** = Optional
Differential pressure and T90-time at different flow rates
P and T90 at flow of
100 200 500 1000 1500 3000 (only /HF) Nl/h
P with new filter element
S-2K 150/F-0,1GF150
0,007 0,011 0,020 0,058 0,135 0,240/0,225 bar
P with new filter element
F-3SS150
0,006 0,012 0,040 0,110 0,215 0,405 bar
T90-time without sample
tube/prefilter
6,0 3,5 1,0 <0,5 <0,5 <0,5 sec.
9 APPLICATION
The M&C gas sample probes type SP2000..., SP2300-H and SP2400-H are used for continuous ex-
traction of gases from dust laden, high-temperature and/or humid processes.