Operator's manual Manual

8
Gas sampling and gas conditioning technology 2-1.0-ME
7 TECHNICAL DATA
compact version SP 2100-H
10S1000
Type SS, stainless steel 316Ti, length 1 m*
max. 600 °C*
0.4 to 2 bar abs.
-20 °C to +60 °C
up to 2 g/m
3
*
120 ml
S-2K 150, filter porosity 2 μm, ceramic
+180 °C self-regulating
after 2 hours
<160 °C, NO
250V-3A AC, 30V-3A DC
1/4"-NPT inside, for max. 10 mm tube connectors
110 up to 240V 50 / 60Hz
start up: 400VA, usual: 100VA, (fuse 6A)
terminals max. 2,5 mm
2
, 2x PG11 cable gland
EN 61010, EN 60335-1
IP54, EN 60529
DN65 PN6,B stainless steel 316Ti
SS 316 / 316Ti, FPM, ceramic
9 kg
not available
P und T90 at flow of:
100
200
500
1000
Nl/h
P pressure loss with new filter element S-2K150:
0,007
0,011
0,020
0,035
bar
T90 time-without sample tube/prefilter-
6,0
3,5
1,0
<0,5
sec
8 APPLICATIONS
The patented M&C SP2100-H gas sample probe is used for continuous gas sampling in processes
with dust densities of up to 2g/m³, operating pressure up to max. 2 bar abs., temperatures of up to a
maximum of 600°C or higher gas humidity. Thanks to its compact design it requires only limited space.
It has to be mounted in a weatherproof position.
9 DESCRIPTION
The sample probes are designed for easy installation, reliable operation and trouble-free maintenance.
Filter elements can be changed without the need for tools and without disconnecting the sample line,
the filter chamber can be cleaned easily and the probe tube can be cleaned without removing the
probe:
these are just a few of the many advantages offered by this probe.
The 2 µm ceramic filter element, which has a large surface, is located in a stainless steel filter
housing.
The gas sample probe is heated up to + 180°C using a special self-regulating heating element. No
thermostat or temperature limiter is necessary.
A separate thermo switch is provided for under-temperature monitoring (< 160°C, NO).