Operator's manual Owner's manual

12
Gas sampling and gas conditioning technology 2-1.0.1-ME
NOTE!
Make sure that the connection is leakproof !
Now place the red silicon heat insulation around the connection.
Mounting of sample tube and adapter flange:
Screw either the probe tube directly into the ¾" inner thread in the flange of the probe with the
¾" flat gasket and tighten.
If the heated probe tube type SP30/35 is used then the probe is to be screwed to their flange
(with welded threaded bolt). First insert the flange seal between the two flanges.
If the probe connection does not correspond to the standard flange connection DN65 PN6,
then the optionally supplied adapter flange should be mounted to the probe in the same way.
Attach the flange seal to the probe connection.
Now insert the process-internal probe section of the complete probe unit into the probe
connection using the screws and nuts supplied.
NOTE!
A preferred mounting position is to have the probe with its sample gas outlet
pointing downwards, although this is not absolutely necessary for perfect
functioning.