Operator's manual Owner's manual

11
Gas sampling and gas conditioning technology 2-1.0.1-ME
If the ambient temperature in the area of the connections is >80°C as a result of radiated heat,
then a radiated-heat deflector must be mounted to protect the probe.
The connection's mounting flange connection should comply with DN65 PN6. If other
connection sizes are required, a special adapter flange /S010 can be supplied as an option.
Before mounting, the probe must be adjusted to the existing operating conditions.
The existing operational parameters are to be checked accordingly prior to commencing mounting
work:
weatherproof mounting position
_____present _____set
Under / over pressure situation
mbar bar
Process temperature
°C, Min. °C, Max.
Dust loading
g/m³
Dust composition - grain size
µm
Gas composition
corrosive toxic explosive
Which parameters should be
measured, e.g. O
2
, CO, SO
2
, NO
X
,..,
Vol.% mg/Nm³ ppm
Required amount of gas
l/h, Min. l/h, Max.
Necessary T
90
time
sec.
13 MOUNTING
M&C SP210-H probes are designed for stationary use and if properly selected and mounted a long
service life and minimum maintenance are guaranteed. It is advisable to mount the probe in an
operational position which is at a 10% inclination to the process.
Checking the filter element:
Turn the U-bolt at the front end of the filter receptacle several times to the left until the retaining
bolt can also be turned sideways.
Remove the filter receptacle at the ring from the probe and check whether the filter element is
screwed on tightly.
Replace the filter receptacle.
Place the retaining bolt in its original position and tighten the U-bolt.
Connection of the heated sample line:
In order to connect the sample line, a threaded tube connector for dimension Ø 6x1mm is
available – other dimensions on request
Push the silicon cap onto the end of the heated sampling tube
Introduce the tube connection piece into the bolted pipe joint and connect.
If a PTFE tube is used as sample line, a carrying bracket must under all circumstances be
inserted in the end of the tube in order to prevent the tube being pressed together.
The temperature-resistant, stainless steel connectors supplied by M&C have a double-blade
ring system to ensure reliable sealing. After tightening the nuts of these connectors by hand,
they should then be tightened exactly 1¼ of a turn using a flat spanner and are then properly
mounted.