User Manual

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.6
Technical Data
Technical specications and illustrations are without
obligation, subject to modications. 03.97/06.06
option:
20S908120S9065 ....
option: sample tube or pre-lter
345
20S5000
Gas Sample Probe SP2000-H320/S
Reset
320°C
9 101112
L'L
7
6
5
8
N
1 2 3
Tu<-30 C
4
Tsoll
Netz/Power
To>+30 C
0-320 C
110
tube ø 6 mm
340
Output
DN4/6
230V,50Hz
(115V,60Hz)
condensate
93F0145
340
650
03B1000 bis 03B5050
heated sample line max. 200 C
DN4/6
Dimensions
Gas Sample Probe SP2000-H320/S
Dimensions in mm
SP2000-H320/S
Part No. 20S5000 (a) 20S5000 (a) + 20S9027
Temperature regulation Thermostat adjustable 50-320 °C,
with over temperature limiter
and low temperature alarm
as contact output alarm point ∆T30 °C,
contact rating 250V 3A~ 0,25A =
with FeCu-Ni thermoelement,
(instead of thermostat)
option: external electronic temperature
controller necessary
e.g. Part No. 01B8350
Probe heating max. 320 °C
Ambient temperature +5 °C to +60 °C** optionally with polyester protective housing -20 °C to +60 °C
Volume of filter chamber 120 ml
Sample pressure 0.4 to 2 bar abs.
Filter element ceramic, type S-2K 150*, filter porosity 2 µm
Condensate vessel glass (optional SS 316Ti, Hastelloy
®
), volume 0.4l (0,15l glass ball filling)
Adapter flange for condensate vessel Hastelloy
®
Peristaltic pump SR25.1G, 230/115V 50/60Hz
Ready for operation after 2h
Connections sample gas outlet / condensate outlet hose fitting DN4/6
Connection test gas inlet tube connection ø 6 mm with blind plug, option: ø 1/4" (a)
Power supply 230V 50Hz, 800W, option: 115V 60Hz (a)
Electrical connection terminals max 2.5mm
2
, 2x PG11 cable glands
Electrical equipment standard EN 61010, EN 60519-1
Degree of protection IP54, EN 60529
Mounting flange DN65 PN6, B, stainless steel 316 / 316Ti, option: 3"ANSI 150lbs RF (a)
Connection sample tube G3/4"i
Material of sample contacting parts stainless steel 316Ti, graphite, ceramic, Hastelloy, glass, FPM, PTFE, PVDF
Weight 17 kg
Options
Part-No.: 20 S 9053 2-way ball valve for shut off to process side /VA320
Part-No.: 20 S 9330 3-way ball valve for shut off to process side /3VA320
Part-No.: 20 S 9044 test gas inlet via check valve 0.7 bar /R
Part-No.: 20 S 9065 and following test gas inlet via check valve 0.7 bar /R
Part-No.: 01 B 8350 electronic temperature controller in wall mounting housing
Part-No.: on request heating of the condensate vessel to max. 180 °C
Part-No.: 20 S 9410 protection case of polyester
* Standard
** In case of higher ambient temperatures use option PT100 (Part No. 20S9025) or thermocouple Fe-CuNi respectively Ni-CrNi (Part No. 20S9027 resp. 20S9028) instead of the thermostate
controller. Then, an additional electronic temperature controller (see data sheet 2-5.1) is necessary.
Part-No. ....(a) = Execution power 115V 60Hz, ange 3"150 lbs, test gas connection ø 1/4".
For further technical data please see data sheet SP2000.