Operator's manual Instruction Manual

8
Gas sampling and gas conditioning technology 9-3.5.1-ME
8 TECHNICAL DATA
Version PMA 20
02A1000 = 230V 50Hz, 02A1000a = 115V 60Hz
selectable for 0-3, 0-10, 0-30 and 0-100 vol.% O
2
, linear
analogue meter with a scale of 0-30 and 0-100% for each selected
range
analogue meter with a scale of 0-30 and 0-100% for each selected
range, digital meter, 3 1/2 digit 9 mm high
for 0-100%O
2
reading, selectivity 0,1vol.% O
2
0-1V DC non-isolated, load > 100 KΩ, for the range of 100 vol.%, and
0-20mA or 4-20mA* for the chosen range, non-isolated, max. load
300Ω
< 3 seconds at 60 Nl/hr air
deviation: analogue = ± 1% of span / digital = ± 0,1 vol.% O
2
analogue = < 1% of span / digital = ± 0,1 vol.% O2
no influence up to 45 °C
The oxygen reading varies in direct proportion to changes of the ba-
rometric pressure.
variation in gas flow between 0-60 Nl/hr air will cause a difference of
< 0,1 vol.% O
2
.
0,01 up to 1 bar g, (PMA 20 required admission pressure for compe-
tent flow rate, no pump inside)
Outlet of analyser must discharge freely into atmosphere.
max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7-70
Nl/hr
-10 °C up to +40 °C, dry gas
fixed at +50 °C
-10 °C up to +45 °C
-20 °C up to +60 °C, relative humidity 0-90% RH
internal power unit for 230VAC standard or 115VAC available (a)*
+/-10%, 40-60Hz, 26,5VA
terminals 2,5 mm
2
; 1x PG11, 1x PG13,5 cable gland
Platinum, Glass, Polypropylene, Stainless Steel 316, FPM, Epoxy
resin
PP hose connectors DN 4/6 for 4 mm i.d. and 6 mm o.d. tube
IP 53 EN 60529 / EN 61010
plastic wall mounting housing / blue/grey
height 220 mm, width 214 mm, length 182 mm / approx. 3 kg
Options
Extra charge for PMA 20 in solvent resistant version with special measuring cell type PMC-1LB.
Material: O-ring: Kalrez, flowmeter: PVDF, tubing and fittings: PVDF/SS316.
Extra charge for PMA 20 in chlorine resistant execution with a special measuring cell type PMC-
1CL2. All gas bearing tubes PTFE/PVDF equipped with purge gas connectors.
Extra charge for qualification test according to TA-Luft + EN 14181 resp. 13. and 17. BImSchV
Extra charge for PMA with combined analogue/digital display.
analogue meter with a scale of 0-30 and 0-100% for each selected range, digital meter, 3 1/2
digit 9 mm high
LCD-indicator for 0-100%O
2
reading, selectivity 0,1vol.% O
2
or digital meter 3 1/2 digit 18 mm
high LCD for 0-100% O
2
reading, selectivity 0,1 vol.% O
2