Operator's manual Instruction Manual

16
Gas sampling and gas conditioning technology 9-3.5.1-ME
Check output signals at 0,0% O
2
:
NOTE!
If a gas mixture is analysed, the single gas components have to be checked
concerning potential cross sensitivity and regarded for zero calibration. (see
chapter 16.1.1 and 16.1.2).
Shut off pressure outlet valve and bottle valve.
Disconnect hose connection from the analyser.
Zero calibration is finished.
NOTE!
After zero calibration the span has to be calibrated too.
Output signal
Measurement
0-1 V
0 V
0-20 mA
0 mA
4-20 mA
4 mA