Operator's manual Owner's manual

M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 10 -
8.2.1 mA output
The mA output is available on the back panel of the
PMA100-L (see fig.8) at the 15-pole Sub-D socket X3. The
following figure shows the configuration of the terminal.
The menu-driven handling of the mA outputs is described in
chapter 12.
8.2.2 In- and output contacts
Fig. 9 shows the configuration of the terminal X4. The
following connections are available at the 25-pole Sub-D
plug:
four binary inputs, In1 to In4, with 12V - 24V, max. 20mA,
one supply power contact, Out +24V and Out 0V, with
24V, max. 20mA,
four binary output contacts, Out1 to Out4, with max. 24V,
400mA,
one alarm contact, Alarm MC and Alarm NO, with 48V,
max. 500mA, 15W, and
one status contact output, Status MC and Status NO,
with 48V, max. 500mA, 15W.
The menu-driven handling of the in- and outputs is
described in chapter 12.
Fig. 8: 15-pole Sub-D socket X3
Fig. 9: 25-pole Sub-D plug X4
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Out 4 MC
Out 3 NO
Out 2 MC
Out 1 NO
Alarm MC
Status MC
Out +24V
IN 1,2,4; GND
IN 3 (+24V)
In 1 (+24V)
In 4 (+24V)
Out 4 NO
Out 3 MC
Out 2 NO
Out 1 MC
Status NO
Out 0V
IN 3; GND
Alarm NO
In 2 (+24V)
mA +
mA -