User Manual
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
7.11
Technical specications and illustrations are without
obligation, subject to modications. 10.03/11.06
1 Upper part of housing
2 Lower part of housing
3 Holding angle
4 Glass filter frit
5 Hydrophobic protective membrane
6 Flat ring of Teflon
7 O-ring FPM
8 O-ring FPM
Dimensions
Dimensions in mm
Technical Data
Examples for application
1 Gas cooler
2 Liquid-stop LB-1
4 Liquid drain with peristaltic pump SR25.1
5 Gas sample pump
6 Seperator and automatic liquid drain ADS-SS
Liquid-stop LB-1PV LB-1SS
Part No. 03F4005 03F4000
Gas flow max. 200 Nl/h
Gas pressure 0,3-2 bar abs.
∆P max. 0,5 bar
0,3 -10 bar abs.
∆P max. 0,5 bar
Differential pressure with clean
protection membrane, medium air, 20°C
50 100 mbar
100 200 Nl/hr
Temperature sample gas max. +80 °C max. +100 °C
Ambient temperature 0 °C to +60 °C
Storage temperature -25 °C to +80 °C
Stagnant space 4 ml
Material of gas bearing parts PVDF, FPM, PTFE, Polyester, glass 1.4571, FPM, PTFE, Polyester, glass
Sample gas / drain connections NPT 1/8"i DIN ISO 228/1
Mounting / weight wall mounting / approx. 0,3 kg