Operating Instructions Universal Filter Series FT-...-H2 Gas sampling and gas conditioning technology 5-2.1.
Dear customer, we have made up this operating manual in such a way that all necessary information about the product can be found and understood quickly and easily. Should you still have any question, please do not hesitate to contact M&C directly or go through your appointed dealer. Respective contact addresses are to be found in the annexe to this operating manual. Please also contact our homepage www.mc-techgroup.com for further information about our products.
Contents 1 2 3 4 5 6 General information....................................................................................................................... 4 Declaration of conformity ............................................................................................................. 4 Safety instructions ........................................................................................................................ 5 Warranty .........................................................
M&C TechGroup Germany GmbH Rehhecke 79 40885 Ratingen Germany Telephone: 02102 / 935 - 0 Fax: 02102 / 935 - 111 E - mail: info@mc-techgroup.com www.mc-techgroup.com 1 GENERAL INFORMATION The product described in this operating manual has been examined before delivery and left our works in perfect condition related to safety regulations. In order to keep this condition and to guarantee a safe operation, it is important to heed the notes and prescriptions made in this operating manual.
3 SAFETY INSTRUCTIONS Please take care of the following basic safety procedures when mounting, starting up or operating this equipment: Read this operating manual before starting up and use of the equipment. The information and warnings given in this operating manual must be heeded. Any work on electrical equipment is only to be carried out by trained specialists as per the regulations currently in force.
5 USED TERMS AND SIGNAL INDICATIONS DANGER! This means that death, severe physical injuries and/or important material damages will occur in case the respective safety measures are not fulfilled. WARNING! This means that death, severe physical injuries and/or important material damages may occur in case the respective safety measures are not fulfilled. This means that minor physical injuries may occur in case the respective safety measures are not fulfilled.
6 INTRODUCTION AND APPLICATION The M&C electrically heated universal filter FT-…-H2 reliably separates off solids, in particular extremely fine particles, occurring in analysis techniques in gas filtration, using very fine, deep-acting filter elements. It is constructed especially for hot gas analysis technique at 180°C. 6.1 SERIAL NUMBERS The nameplates bearing the serial number are located on the mounting plate. NOTE! 6.
7 TECHNICAL DATA Filter series FT® Part number Filter finess/material of filter element Power supply Electrical connection Protection classification/electrical standard Ambient temperature max. Sample temperature max. Ready for operation Temperature alarm contact Regulation of temperature Sample connections Stagnant space Filter surface Ma.
8 DESCRIPTION The universal-filter out of PTFE with integrated filter elements (fineness 3 or 0,1 µm) is placed in an electrically heated 2-part aluminium construction. The temperature is controlled by an integrated compact capillary sensor thermostat with an excess temperature limit switch with reset key at +30°C above Tset and an alarm function for low temperature alarm at -30°C below Tset with voltage free status-alarm contact.
9 RECEPTION AND STORAGE The universal filter series FT-…-H2 is normally delivered in one packaging unit: The universal filter series FT-…-H2 should be removed carefully from the packaging and checked immediately for completeness against the delivery note. Check the goods for any damage incurred during transport and if necessary inform your transport insurer of any damage.
12 ELECTRICAL CONNECTION WARNING! When connecting the equipment, please ensure that the supply voltage is identical with the information provided on the model type plate.
Figure 2 Electrical connecting diagram Gas sampling and gas conditioning technology 5-2.1.
13 STARTING Before using the equipment for the first time, check that the safety measures specific to the installation and process are complied with! For the carried media the corresponding safety regulations and measures have to be considered. WARNING! Before starting up check whether the mains power supply voltage corresponds with the information stated on the nameplate. Beware of touching the filter surface during operation. Serious burns can occur because of high surface temperatures.
14 MAINTENANCE The safety instructions specific to the plant and process are to be consulted prior to any maintenance work! WARNING! Aggressive condensate is possible.
15 SPARE PARTS LIST Wear, tear and replacement part requirements depend on specific operating conditions. The recommended quantities are based on experience and they are not binding. Universal filter series FT®-..-H2 (C) Consumables, (R) recommended spare parts, and (S) spare parts V/E/T Recommended quantity For operating period [years] 1 2 3 90 F 0005 Filterelement F-3G, Glas, 3µm, 75mm C a.t.r. a.t.r. a.t.r. 90 F 0016 Filterelement F-0,1GF, Glasfaser, 0,1µm, 75mm C a.t.r. a.t.r. a.t.