6 INTRODUCTION
Hot sample gas very often contains solids like dust, taar or grime and water vapour. These materials
cause problems for a continuous gas analysis, especially when the gas temperature falls below the
dew point. The sample line will be blocked and the analyser will be damaged. Therefore the solids
have to be separated and a temperature drop below the dew point has to be prevented. For this
complex of problems the electrically heated M&C universal filter type FSS-../H350 was designed.
7 APPLICATION
The electrically heated M&C universal filter type FSS../H350 separates fine dust particles of 2 or 3 µm
or solids like taar or grime from the sample gas at operating temperatures up to 350 °C. For an
efficient filtration very fine filter elements with high deep action are used. For separation of e.g. taar
and grime a special filter cartridge filled up with spun glass is used.
8 TECHNICAL DATA
FSS-2K/H350
FSS-3SS/H350
FSS-FW/H350
Version 115V AC
01F3035
01F3030
01F3045
01F30..-a
ceramic
stainless steel
SS316
spun glass
2 μm
3 μm
140 cm
2
5 bar g
-20 °C to +60 °C
130 cm
3
Swagelok tube connector ø6 mm, optional ø1/4"
incorporated capillary sensor thermostat, with
temperature limiter switch and low temperature alarm
contact
adjustable from +50 °C to 350 °C, set in factory 250 °C
alarm point at -30 °C below TSet change over contact,
voltage free, contact rating 250V, 3A ~ 0, 25A =
230V 50Hz, 800VA
115V 60Hz
Terminals 4 mm
2
, 2x cable glands PG13
Wall mounting
IP 54 EN60529 / EN61010, EN60519-1
250 x 300 x 300 mm (w x h x d) / 11 kg
filter chamber: stainless steel 316Ti / sealing: graphite