Manual
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
6.7
Technical specications and illustrations are without
obligation, subject to modications. 09.96/06.06
Aerosol-Filter CLF-5
External mounting in the outlet of an ultra low gas cooler EC-30 when sample tends to form aerosols. Part Number 03F3000
Technical data see data sheet 5-1.10.7 Fluid particle filter CLF-5
* Maximum values in technical data must be rated in consideration of total cooling capacity at 25 °C ambient temperature.
Cooler series EC
®
version EC-30 version EC-30/FD
Part No.: 02K6000 (a)** 02K6010 (a)**
With integrated universal unit: EC-F to force the ventilation EC-FD ventilation and condensate removal
Gas connection tube connector DN 4/6 i./o. d. mm
Condensate connections 3x tube connector GL25-12 mm 1x tube connector DN 4/6 i./o. d. mm
Material of sample contacting parts Duran Glass, PTFE, PVDF Duran Glass, PTFE, PVDF, Novoprene
Single stream, gas flow rate 90 Nl/hr-250 Nl/hr*
Gas pressure max. 3 bar abs.
Ambient temperature +5 to +45 °C
Storage temperature -20 to +60°C
Sample outlet dew point -30 °C
Sample inlet temperature max. 180 °C*
Sample inlet dew point max. 80 °C*
Cooling capacity at 25 °C ambient max. 860 kJ/hr*
Main power connection / Power consumption 230V 50Hz 380VA or **Part No. ...-a = 115V 60Hz, 380VA
Start up time < 60 min.
Stagnant space approx. 160 ml
∆P at 250 Nl/hr flow rate 5 mbar
Electrical connection 2.5 mm
2
terminals, 2x PG13,5 cable glands
Status alarm 2 changeover contacts, potential free contact rating 250V, 3A, 500VA, 50W
Case protection / Electrical equipment standard IP20 EN 60529 / EN 61010
Method of mounting / Case colour 19" rack or wall / case colour RAL 9003
Dimension / Weight 84 HP x 8U x 360 mm / 37 kg 84HP x 10U x 360 mm / 42 kg
Option
Technical Data