Operator's manual
18
Gas sampling and gas conditioning technology 2-1.1.7.6-ME
11 RECEIPT AND STORAGE
Carefully remove the dilution unit and any accessories from the transport packaging immediately af-
ter receipt and check the completeness of the delivery against the packing list.
Check the goods for possible transport damages and, if necessary, notify immediately your trans-
port insurer of any damage.
NOTE!
The dilution unit should be stored in a protected, frost-free room!
12 INSTALLATION INFORMATION
The safety rules and regulations for the prevention of accidents must be observed during installation
and also subsequent operation. The information in chapter 3 “Important Safety Information“ must be
observed.
The following also applies:
Select the optimal sampling point according to the generally applicable directives or coordinate
with the responsible departments.
NOTE!
The dilution unit must be checked for its suitability for use with the available
operating parameters prior to installation (see type plate).
13 INSTALLATION
The M&C dilution units DIL-1... are designed for stationary use. With correct selection of the sample
point and proper installation, they will work for many years with a minimum of maintenance required.
13.1 CONNECTION OF THE SUPPLY OR SAMPLE LINES
NOTE!
Increases of pressure due to use of long sample gas lines at the outlet may
have a great influence on the measuring result. For this reason, we recom-
mend the following nominal widths and lengths for the sample lines in de-
pendence on the type of injector:
Injector type I:
i
6mm max. 50m
i
8mm max. 150m
Injector type II:
i
8mm max. 15m
i
10mm max. 40m
i
12mm max. 80m