Operator's manual

13
Gas sampling and gas conditioning technology 2-1.1.7.6-ME
8 TECHNICAL DATA
Dilution ratios of the critical orifices „a“ to „g“
a
b
c
d
e
f
g
500
200
100
50
30*
20
10:1
Sample flow rate with critical orifices „a“ to „g“
1,4
2,7
5,5
11
19*
28
55l/h
1)
Possibility to adapt the dilution factor
With dilution gas pressure adjustment -5% to +30%
2)
Dilution gas flow rate with injector version I or II
I: 480 - 600 Nl/h, optional II: 1800 3000 Nl/h
Dilution gas pressure on inlet of pressure controller
Min. 4,5 bar, max. 16 bar
Bypass injector /B: pressure / Gas consumption / flow
rate
approx. 2 bar : injector gas approx. 300 l/h : sample
gas approx. 150 l/h
Process pressure
0,9 to 2 bar abs. at constant pressure
Fault caused by process temperature variations
No fault, operation independent of process tempera-
ture
Fault caused by process low or over pressure
Influence negligible at pressure changes 200mbar
and if test gas is given to the probe under process
conditions. Otherwise proportional to pressure
changes.
Fault caused by atmospheric pressure variations
< 1% with a variation of 50 mbar
Material in contact with the sample gas
Stainl. Steel 316Ti, quartz glass, FPM, graphite
Power supply for DIL-1/H
230V 50Hz, (optional 115V 60Hz), 800W
Temperature controller for DIL-1/H
Capillary thermostat adjustable 0-180 °C, with ex-
cess temperature limiter and low temperature alarm
as changeover contact, potentialfree alarm point T30
°C to T
Set
,
contact rating max. 250V 3A~ 0,25A =
Weight
approx. 8 kg
* Standard, others to be indicated along with order, intermediate values possible.
1)
approx at 3 bar dilution gas.
2)
-5%, however not possible with orifice „g“.
Type
M&C Dilution unit DIL-1/(H) with orifice „e“, dilution ratio 30:1, Standard
DIL-1
Dilution unit non heated
DIL-1/H(a)
Dilution unit electrically heated to 180°C, (a) = voltage 115V 60Hz
Option I
Bypass injector B
Option II
Additional sample gas outlet undiluted
DIL-1/..-A
Option: mounting set for dilution unit, 1 pressure controller and 2 manometers
DIL-1/..-A1
Option: mounting set for dilution unit/Bypass injector, 2 pressure controllers and 3 manometers
DIL-1/..-S
Option: control panel with 1 pressure controller, 2 manometers, flowmeter, 2 shut-off valves
DIL-1/..-S1
Option: control panel with 2 pressure controller, 3 manometers, flowmeter, 2 shut-off valves