User Manual

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
12.1
Technical specications and illustrations are without
obligation, subject to modications. 07.00/12.06
Technical Data
Dimensions CG-2, CG-2M, 19-rack
housing unheated gas inlet and outlet
Functional diagram type CG-2M
Versions CG-2, CG-2H-W without Y1/2, Version CG-2MH-W without Y2
Dimensions in mm
132.5
7
10
325
D C
A
B
G1/4"i
view from the backside
B1.0
B1.2
B1.1
40260
57
483
129
435
400
400
(3HE)
(84TE)
B 1.0
B 1.1
B 1.2
350
350
E1
Y1
Y2
B
A
C
D
E2
E2
A Status alarm temperature
B Power supply
B 1.0 Temperature control
B 1.1 Temperature alarm
B 1.2 Temperature display
C Sample gas inlet in
D Sample gas outlet
E 1 Tube furnace
Y 1 3/2-way solenoid valve (only in version CG-2M, CG-2MH-W)
Y 2 3/2-way solenoid valve (only in version CG-2M)
E 2 Heating sample gas inlet, outlet and Y1 (only in version
CG-2MH-W) or heating sample gas inlet and outlet (only in
version CG-2H-W)
265
2
437
423
Temp. controller
Power
External
control/
Alarm
Sample gas outlet
DN 4/6*
DN4/6*
Sample gas in
Cabinet
mounting plate
Sample gas inlet
DN4/6
converted
3 5 0.0
3 5 0.0
457
245
145
by customers
Dimensions CG-2H-W, CG-2MH-W, wall mount-
ing housing, heated gas inlet and outlet
Gas Converter Series CG® Version CG-2 Version CG-2M Version CG-2H-W Version CG-2MH-W
Part No. without catalyst cartridge 50A1600 (a) 50A1500 (a) 50A1900 (a) 50A1920 (a)
Part No. catalyst cartridge type C, carbon-molybdenum 95A9003
Part No. catalyst cartridge type SS, metal filling 95A9004
Gas inlet and outlet non heated (cold) or heated (hot) non heated heated
Housing 19" rack mounting 3U depth 350 mm wall mounting housing
Bypass solenoid valves no 2x no 1x
Temperature of sample gas max. +80 °C, dew point dry max. +250 °C, dew point<160 °C
Temperature range, adjustable max. 680 °C
Gas flow rate max. 90 Nl/hr
Operating pressure max. 2 bar abs.
Warm-up time approx. 30 min.
Operating temperature at N0
2
/ NO converter with type C 350 °C
Operating temperature at N0
2
/ NO converter with type SS 660 °C
Conversion N0
2
/ N0 Efficiency >95 % with new catalyst
Live of catalyst N0
2
/ N0 to renew approx. every 6 months as preventive measure
Differential pressure <20 mbar
Ambient temperature +10 °C to +50 °C
Storage temperature -25 °C to +65 °C
Relative humidity <80 %
Sample gas connections PVDF fitting G1/4” i DIN 228/1 ø6 mm tube connector, SS316Ti*
Power supply 230V 48-62Hz or (a) 115V 48-62Hz
Power consumption 520W 620W 940W
Electrical connection Main power plug connector incl. two fine fuse 5 x 20 mm,
230V: 6,3A, 115V: 10A, 2 m cable and shock-proof plug.
Alarm-/control signals 9-pin sub D connector
Status signals for temperature NO contact-potential free, contact rating max. 24V, 1A
Materials of sample contacting parts Stainless steel SS 316Ti, PTFE, FPM Stainless steel SS 316Ti,
alu hardcoated, FPM
Degree of protection IP 20 EN60529
Weight approx. 6 kg
Electrical equipment standard EN 61010, EN 60519-1
Option for 19”-rack housing Wall mounting bracket, 3U-84HP, Part No.: 50A3000
* Standard ø 6 mm, for ø 1/4" – please indicate with order. Other executions on request