Operator's manual
20
Gas sampling and conditioning technology 10-1.1-ME
15.5 ADSORBER CARTRIDGE REPLACEMENT
The following steps must be carried out:
Relieve the pressure in the combustion air conditioning unit; switch off the “Air/On” switch
internally or externally (LED goes out).
Remove the screw cap by turning counter-clockwise.
Remove the adsorber cartridge from the holder by turning the stirrup bolt.
Refitting takes place in reverse order.
NOTE!
In order to ensure the necessary tightness, it is important to ensure
that the cartridge is inserted fully into the holder !
15.6 O-RING SEAL REPLACEMENT
With each cartridge replacement, it is recommended to also replace the adapter seals.
W A R N I N G !
Hot catalyst cartridges. Contact can result in serious burns. Wear
protective gloves and secure cartridge against unauthorised
access !
Figure 4 shows the position of both inner and both outer O-ring seals.
Cartridge adapter
Inner seals
Outer seals
Figure 4 Catalyst cartridge adapter