Model 1134 AccuRead Blood Pressure Monitor & Clock User Manual Distributed by: The Lumiscope Co., Inc www.lumiscope.
2 Table of Contents Introduction .................................................................... 3 How This Monitor Works.................................................. 4 About Blood Pressure ..................................................... 5 Parts Identification ......................................................... 6 Unit Display Symbols ...................................................... 7 Unit Display Explanation ................................................. 8 Features .........
How the 1134 Works About Blood Pressure 5 Your model 1134 uses the oscillometric method to detect What is Blood Pressure? your blood pressure. Before the cuff starts inflating, Blood pressure is the pressure exerted on the artery the device will establish a baseline cuff pressure, which tube while blood flows through the arteries. The pressure is equivalent to the air pressure.
6 Parts Identification 7 Unit Display Symbols Display Symbols Date/ Time KPa AVG Systolic Pressure 1 2 3 Start/Power Pre-formed Cuff Memory Zone Select Air Tube Average Selection USB Plug /PC Link Memory Select Cuff Plug Outlet WHO Indicator AC Adaptor Outlet 4 Diastolic Pressure Pulse Value
Unit Display Explanation 8 Display Explanations User 1-4: Appears when the unit is being operated by User 1 through User 4. Low Battery Indicator: Appears when the batteries are low or need replacement. Pulse Symbol: Shows the pulse rate per minute. Features Functional • One touch operation • Large LCD display • Irregular Heartbeat Detector: This unit can detect a disturbance in the regular rhythm of the heartbeat, which may cause or indicate cardiovascular disease.
10 Installing the Batteries 1. To insert the batteries A. Press down on the latch and lift the cover towards you to open the battery compartment. B. Insert 4 "AA" batteries as indicated. Make sure to match polarities (+, –) on the batteries. C. Replace the battery cover by inserting the two hooks at the bottom into the slot, then push on the top area. Note: Please remove batteries when not in use for an extended period of time. Do not dispose batteries with household garbage. 2.
12 Applying the Cuff To apply the cuff A. Plug the end of the tubing on the cuff into the connector at the base of the monitor. 2~3 cm (0.8” ~ 1.2”) Main Artery Left Arm B. Remove any clothing that restricts the upper left arm. Wrap the cuff around your upper left arm. The rubber tube should point in the direction of your hand. The bottom edge of the cuff should be 0.8" ~ 1.2" (2~3 cm) above your elbow with the tube positioned over the middle of your arm. 13 Applying the Cuff D.
14 Measurement Procedures WHO Lighted Indicator 15 To start the measurement Showing levels of Blood Pressure Reading Note: Standards for assessing high blood pressure, without regard to age, have been established by the World Health Organization (WHO), as shown in the chart as below. This unit offers 6-color light signs after user's measuring. From the light sign, it will help users to find out their health situation. Different levels show the corresponding light color.
16 Irregular Heartbeat Dectection An irregular heartbeat may cause or indicate cardiovascular disease. This unit has a unique function that alerts the user to the presence of an irregular heartbeat. Heartbeat Readings 17 Regular Heartbeat reading but below normal When taking your blood pressure reading, if your reading displays a value below 60 beats/sec, then the unit will display the Pulse Value and "LO" symbols alternately.
18 Memory Recall To see the previously measured values A. Push the "PAGE" button to select one of the four Memory Zones. B. Once one of the four Memory Zones has been selected, push the "MEMO" button twice. The first blood pressure reading will be displayed. C. To view additional readings, push the "MEMO" button successively to go through the various readings. (1~99) 19 Average Recall This unique monitor automatically offers a set of five average values: (Hour, Day, Week, Month and All).
20 Clock Settings 1. To set time and date A. Press the " " button and hold for at least 5 seconds. You can start setting the year when the "YEAR" digit blinks. B. Then press " MEMO " to increase the year and "PAGE " to decrease the year. C. After finishing "YEAR" setting, press " " button again. Now you can adjust Month/Date/Hour and Minute digits as you desire. 2. How to set Alarm Clock A. Press " " button and hold for at least 5 seconds. The "YEAR" digit blinks.
22 Maintaining & Storing Unit A. Use a dry, soft cloth to clean the unit. If desired, use a cloth lightly dampened with tap water. B. Do not use alcohol, benzene thinner or other harsh chemicals to clean the device or cuff. Problem Nothing is displayed when the "ON/OFF" key is pressed. C. Before measuring, please wash your hands. Do not wash or wet the arm cuff. D. When storing this unit, keep it away from direct sunlight, high humidity or dust.
24 Specifications Two Year Limited Warranty 25 Model : 1134 Power source : 4 "AA" Alkaline batteries or AC/DC adapter Measurement method : Oscillometric Measurement range : Pressure: 20~280 mmHg Pulse: 40~195 beats/minute Accuracy : Pressure: ±3 mmHg Pulse: within ±5% of reading Pressure sensor : Semi-conductor Inflation : Pump driven purchase.